DEPARTMENT OF HEALTH AND HUMAN SERVICES
PUBLIC HEALTH SERVICE FOOD AND DRUG ADMINISTRATION
FIELD TEST SERIAL NO. (1-8)
HEAD AND NECK RADIOGRAPHIC SYSTEMS FIELD TEST RECORD (Use Form FDA 2782, Field Test Record Continuation, if more space is needed.)
Card No. (9-10) Test Procedure 1. HN 11 Component Certification Information 2. Indicate the status of each as follows: C - Certified
Print Legibly. Use Black Ball Point Pen. Enter One Character Per Box.
HN
REGIONAL REVIEW (NAME)
V - Certified with a Variance X - Not Present or Not Applicable
13
N - Not Certified
Beam Limiting Device
High Voltage Generator
Cassette Holder
14
Tube Housing Assembly
15 18
X-ray Control Tube Housing Assembly With Beam Limiting Device
16
Other (Specify)
17 20
19
10
3. Indicated Source To Image Distance (SID)
q
inches
OR
q
cm
21
23
24
27
Test Setup
S6
Technique Factors 4. Timer mode of operation during testing
5.
kVp
29
6.
31
mA
28
M-manually set time or mAs P- phototimer MDH (Pulse Exposure) 4.5 mm Al Beam Quality 9.
32
7.
q
34
sec OR pulses
35
8.
mAs
38 44
39
41
42
q
mR
@
4.5 mm Al
45
10.
q
49
mR @ 3.5 mm Al
50
11.
q
54
mR @ 2.5 mm Al
55
12.
q
59
mR @ 1.5 mm Al
60
13. Techinque Factors indicated Before Exposure
64
14. Exposure Terminated After Preset Time Interval, Preset mAs, Or Preset Radiation Exposure To Image Receptor
11
15. Warning Label Present Y-YES N-NO
Y-YES N-NO
12
Y-YES N-NO
Reproducibility 17.
q
13
mR
18.
msec
15
16. MDH Threshould Setting, 0.53-3 phase, 0.2-1 phase 19.
q
19
mR
20
20.
23
msec
24
21.
q
28
mR
29
22.
32
msec
11
33
0
q
37
q
38
mR
41
msec
46
23.
24.
42
46
47
50
PAGE 1 of 3 PAGES
Created by: PSC Media Arts (301) 443-2454
FORM FDA 3297 (4/84)
EF
DEPARTMENT OF HEALTH AND HUMAN SERVICES
PUBLIC HEALTH SERVICE FOOD AND DRUG ADMINISTRATION
FIELD TEST SERIAL NO. (1-8)
HEAD AND NECK RADIOGRAPHIC SYSTEMS FIELD TEST RECORD (Use Form FDA 2782, Field Test Record Continuation, if more space is needed.)
Reproducibility (Continued) 25.
q
Print Legibly. Use Black Ball Point Pen. Enter One Character Per Box.
HN
REGIONAL REVIEW (NAME)
mR
26. Data Here If Any Two Of Items 17, 19, 21, And 23 Differ By More Than 10 Percent Of Largest Value
msec
51
27.
55
q
56
28.
59
msec
11
29.
mR
60
q
64
mR
65
30.
68
msec
69
31.
q
73
mR
74
32.
77
msec
11
33.
q
15
mR
16
34.
19
msec
20
35.
q
24
mR
25
36.
28
msec
29
33
34
37
Linearity 37.
mA
38
40
42. Distance from Base of Test Stand To Face of Cassette Holder
q
cm
12
61
43. Distance from Face Cassette Holder to Image Receptor
64
If Change in mA Caues a kVp Shift, Readjust kVp Setting to Value Selected at Item 5 Above 38.
q
q
cm
65
mR
67
41
39.
q
45
mR
44. Outside Separation of Image of Focal Spot Strips
q
cm
68
70
46
40.
q
50
mR
51
41.
q
55
mR
56
60
X-Ray Field/Indicated Field Size Comparison (Fixed Collimation Only) Indicated Field Size Horizontal Dimension
45.
q
inches
OR
q
cm
11
46. Indicated Field Size Vertical Dimension
q
13
inches
14
q
16
cm
17
47. X-Ray Field Image Horizontal Dimension
q
19
cm
20
48. X-Ray Field Image Vertical Dimension
q
22
cm
23
25
26
28
Actual Versus Indicated Field Size
13
49. Beam Limiting Device Numerically Indicates Field Size Indicated Field Size Horizontal Dimension Y-YES N-NO
29
inches OR cm
50.
q
q
30
51. Indicated Field Vertical Dimension
q
32
inches OR
33
q
35
cm
36
52. Light Field Horizontal Dimension
q
38
cm
39
Light Field Vertical Dimension
q
41
cm
53.
42
44
45
47
FORM FDA 3297 (4/84)
PAGE 2 of 3 PAGES
DEPARTMENT OF HEALTH AND HUMAN SERVICES
PUBLIC HEALTH SERVICE FOOD AND DRUG ADMINISTRATION
FIELD TEST SERIAL NO. (1-8)
HEAD AND NECK RADIOGRAPHIC SYSTEMS FIELD TEST RECORD (Use Form FDA 2782, Field Test Record Continuation, if more space is needed.)
Illuminance 54.
total ambient
Print Legibly. Use Black Ball Point Pen. Enter One Character Per Box.
HN
REGIONAL REVIEW (NAME)
=
q
55.
fc
=
q
fc
48
50
57.
q
total
ambient
51
53
56.
total ambient
=
fc
=
q
fc
54
56
total
ambient
57
59
13
X-Ray Field/Light Alignment and Size Comparison 58. X-Ray Field Horizontal Dimension 60. Light Field Horizontal Dimension 62. Horizontal Misalignment
cm
q
60
q
62
cm
59. X-Ray Field Vertical Dimension 61. Light Field Vertical Dimension 63. Vertical Misalignment
q
cm
63
q
65
cm
66
q
68
cm
69
q
71
cm
72
X-Ray Field/Image Receptor Centers Comparison 64. Centers Misalignment
73
74
75
q
cm
11
PBL X-Ray Field/Image Receptor Size Comparison 65. Horizontal Film Dimension 66. Vertical Film Dimension 67. Indicated
SID
q
12
Inches
OR
q
cm
68. Type of Positive Beam Limitation (PBL) A. Automatically Adjusts X-ray Field B. Prevents Production of X-rays Until Manual Adjustments are Made
13
q
15
Inches OR
16
q
18
cm
19
q
21
Inches OR
22
q
24
cm
25
27
28
31
Y-YES N-NO
32
69. Is the PBL Currently Operating in Conformance with its Design
33
14
70. Distance from Film Cassette to Plastic Cassette
q
cm
34
71. Light Field Horizontal Dimension 73. Indicated SID
q
37
q
cm
38
q
40
Inches
72. Light Field Vertical Dimension
OR
q
cm
41
cm
43
44
74. Light Field Horizontal Dimension PBL Operation
46
q
47
cm
50
q
51
53
75. Light Field Vertical Dimension
cm
54
56
76. In PBL Mode, Adjustment Possible To Field Size Smaller Than Image Receptor
57
Y-YES N-NO
77. Automatic Return To PBL When Image Receptor is Changed
58 59
Y-YES N-NO
Y-YES N-NO
78. X-ray Production Prevented At SID's Where Operation Is Not Intended REMARKS
FORM FDA 3297 (4/84)
PAGE 3 of 3 PAGES